Investigations Into X-ray Fluorescence Analysis of Slurry Streams at Mount Isa

A commercial Philips X-ray Fluorescence (X.R.F.) analyser was converted to an experimental instrument suitable for determination of the accuracy to be expected of X.R.F. on line analyses of concentrator slurries. Problems with stability of slurry sample presentation were eliminated by a number of minor modifications.By the use of multiple linear regression techniques instrument calibration equations including error terms were determined for copper flotation feed slurries. Results indicated that the accuracy of copper and iron analyses of flowing slurry samples could approach that of static pressed powder samples.Possible means of reducing the error terms to a level of at least that of pressed powdel' samples involve alternative methods of compensating for the per cent solids in the slurry, Design is under way for installation of a plant unit.